The Nd: Yag Laser


Nd:YAG Layout | Operations Data | Nd:YAG Statistics | Gun Hutch

 

CW phasemeter ( DC - 10 Hz bandwidth ) Shown over 48 hours

Current Status Capabilites

 - Photocathode and CO2 slicing fully available on-demand

  - Electron beam-synchronized optical pulses available for users

10 mJ, 14 ps @ 1064 nm in laser lab ( exclusive of slicing )

100 mJ, 10 ps @ 532 nm in laser lab or FEL room

50 mJ, 8 ps @ 266 nm in gun hutch and laser lab

  - Exceptional availability: deliver light ~130 days/year, for a total 1500 running hours.

  - High reliability: system typically ready for gun operation within minutes, including daily performance characterization.

Demonstrated ATF Laser Performance

Energy (dual pulse mode)  
UV on Cathode 0-50 mJ
IR at CO2 table 7mJ
Laser output:total IR 30mJ
    IR into 2w 5mJ/ pulse
    Green 1mJ/pulse
    UV 200 mJ
   
Repetition rate 1.5, 3 Hz
   
Pulse Duration (FWHW):  
    Oscillator IR 7ps
    Amplified IR 14 ps
    Green 10 ps
    UV 8 ps
   
Beam on cathode(FWHW): 0.2 -  3mm
   
Top-Hat Beam Profile Modulation (P-P) <20%
   
Shot-to-shot stability (rms):  
    Timing <0.2 ps
    Energy 2%
Pointing (Fraction of Beam) <0.3%
   
Drift (8 hour P-P)  
    Timing 1ps
    Energy <15%
    Pointing (fraction of beam) <1%

 

Time-Bandwidth Products, Inc. Nd:YVO4 Oscillator

  - Improved stability in both energy, phase, and pointing

  - High reliability reduces maintenance - realignment should be unnecessary

  - Old oscillator provides emergency backup in case of failure

 

  LWE-131 GE-100
Pulse duration (FWHM) 20 ps 7 ps
CW power ~65 mW (currently) 500 mW
Amplitude stability <0.3% (10kHZ-10 mHz) <0.2%
Pointing stability < 50 microradian <10 microradian
Phase stability

<1 ps p-p (minutes)

< 1ps/hour drift

<0.3 ps p-p (minutes)

< 0.5ps/hour drift

Estimated laserdiode lifetime > 30,000 hours* >10,000

*-has operated 36,000 hours

Diagnostics

Continuously on-line:

  - Scanning autocorrelator for oscillator monitoring

  - 14 CCD cameras for transverse mode and position measurements with beam profile analyzer

  - 5 pyroelectric energy measuring probes with pickoffs and calibrated readout*

  - 7 High speed silicon photodiodes

  - Laser-RF Phasemeter*

  - 16-channel thermocouple temperature monitoring

Additional

  - 2 ps resolution streak camera (Instrumentation Div.)

  - CW laser power meter

  - 2 GHz digital sampling oscilloscope

* - output available facility-wide

 

Top of Page

Last Modified: December 3, 2007
Please forward all questions about this site to: Vitaly Yakimenko